Sodium- Determination by AAS
COEI-2-SODIUM Determination of sodium by absorption atomic spectrometry
- Principle
The sodium is determined after mineralisation by dry process by atomic absorption spectrometry.
The addition of a spectral buffer (cesium chloride) to avoid ionisation of sodium is necessary.
- Apparatus
2.1. Glassware
- Graduated flasks 50 and 100 ml (class A)
- Graduated pipettes 2.0; 5.0; 10.0 ml (class A)
- Automatic pipette 1000 μl
- Cylindrical vase 100 ml.
- Instrumental parameters: (given as an example)
- Atomic absorption spectrophotometer
- oxidant air-acetylene flame (rate-air: 3.1 l/mn; rate-acetylene: 1.8 l/mn)
- wave length: 589.0 nm
- hollow-cathode lamp (sodium)
- width of slit: 0.2 nm
- intensity of the lamp: 5 mA
- no correction of non specific absorption
- Reagents
3.1. Pure demineralised water for analysis
3.2. Pure nitric acid for analysis at 65%
3.3. Cesium chloride solution at 5% in cesium: Dissolve 6.330 g of cesium chloride in 100 ml of pure demineralised water for analysis.
3.4. Sodium reference solution at 1 g/l commercial or prepared as follows: dissolve 3.6968 g NaNO3 in water, adjust at 1 l.
3.5. Diluted sodium solution at 10 mg/l:
Place 1 ml of the reference solution at 1 g/l in a 100 ml graduated flask, 1 ml of nitric acid at 65%, complete to volume with pure demineralised water for analysis.
3.6 Set of calibration solutions 0; 0.25; 0.50; 0.75; 1.00 mg of sodium per litre:
In a series of 100 ml graduated flasks, place 0; 2.5; 5.0; 7.5; 10 ml of the diluted sodium solution; in all the graduated flasks add 2 ml of the cesium chloride solution and adjust the volume at 100 ml with pure demineralised water for analysis.
The calibration solutions prepared contain 1 g of cesium per litre; they are stored in polyethylene flasks.
- Preparation of samples
4.1. Liquid or solution oenological products
In a 50 ml graduated flask, place 1 ml of the cesium chloride solution at 5% and a volume of sample after having been completed to volume with demineralised water, the concentration of sodium to be measured is below at 1 mg/l.
4.2. Solid oenological products
Proceed with a mineralisation by dry process (take up the cinders in 2 ml of hydrochloric acid in a 100 ml flask, add 2 ml of cesium chloride at 5% and complete to volume with demineralised water).
Perform a blank test with demineralised water.
- Determinations
Present successively calibration solutions.
Perform an absorbance reading for 10 seconds; perform two measurements.
Set up the calibration curve (absorbance depending on the concentration in mg/l of sodium).
Then present the samples; determine the concentration of sodium of the diluted samples in mg/l.
Calculate the concentration of sodium in the oenological products in mg/kg.
The dosages of air-acetylene flame are performed manually.